1-ID Insertion Device Information
The insertion device that is currently installed on 1-ID is a standard APS
Undulator A. The following information is availiable.
- How to scan the
undulator.
- Actual harmonic
energies.
- Undulator A Tuning Curves
- ANL/APS/TB-3,
"Undulator A Characteristics and Specifications" B. Lai, A. Khounsary, R.
Savoy, L. Moog, and E. Gluskin (February 1993)
- ANL/APS/TB-17, "Undulator A Characteristics and Specifications: Enhanced
Capabilities", R. J. Dejus, B. Lai, E. R. Moog, and E. Gluskin (May, 1994).
Dean Haeffner ([email protected]) / 3/26/99
EPICS 3.13.1 compatible instructions
- Insert the PV "1id:ID01_energy" (1id:ID01_gap to scan the gap) into the
"Drive" field of the MEDM scan tool. This PV can be found in the xxIDCtrl.adl
screen.
- Insert the PV "ID01:Energy" into the "Read" field of the MEDM scan tool.
- Set the "Gap Deadband" on the undulator control screen to the desired
value (1 micron approximately corresponds to 1 eV in 1st harmonic energy.)
There is no advantage to setting the deadband to greater than 3 micron and
lower settings can considerably slow down the undulator motion.
- If you want the energy to read for a particular harmonic, set the
"Harmonic Select" field on the undulator control screen.
- Enter the desired scan range and step size (or number of steps) and start
the scan.
Important Note: Always scan the undulator from large
to small gap. The mechanical supports on the undulator are subject to
tremendous forces and has some amount of hysteresis. The encoders are calibrated
for closing, not opening, the gap. Experimental results have been measured for
the Kohzu DCM set to 19.6 keV using both the third and fifth undulator
harmonic. Note that these values were for UA#7 prior to the Sept. 98 collision
with the storage ring. They will be remeasured and updated as soon as possible.
The energy positions given by the undulator control screen for undulator
harmonics are not accurate. The figure below shows the differences between
measured and calculated values on for the 1-ID undulator.
The following is the calculated tuning curve for a standard 3.3-cm period
Undulator A.
The following is a measured tuning curve from 1-ID. The monochromator had Si
(111) crystals. Note that the data is "raw" and does not include ion chamber
efficiencies.